Figure 17: AllCall© Bed Station Schematic
AllCall© Bath Station
The AllCall© Bath Station, is very similar to the
Bed Station in that it uses the same plastic enclosure
and wireless transmitter. At that point, however the
similarity ends. As can be seen in the illustration, the
Bath Station is activated by pulling the string. When
the slide switch moves to the alarm position (down),
the red LED comes on and shows the call station to
be in alarm. Once the resident’s need is handled, the
slide switch is then firmly pushed back to the reset
position (up).
The slide switch is single pull, double throw and it is mounted in a
Tektone Faceplate. One side of the switch controls the wireless transmitter,
and the other side controls the LED. On the inside of the Call Station box,
there are two primary components, a wireless transmitter and a LED
battery board. The transmitter controls the communication with the control
This area is reserved for Figure 17